Ask Latest Price
Video Channel
Brand Name :AI
Model Number :AOI
Certification :ISO
Place of Origin :CHINA
MOQ :1 SET Minimum Oxygen Index Test Apparatus
Price :Negotiable
Payment Terms :Western Union, T/T, L/C
Supply Ability :200 sets per month
Delivery Time :40 days
Packaging Details :Minimum Oxygen Index Test Apparatus packed in Plywood case
Test standard :ASTM D2863,ISO 4589-2,NES 714
Size :400*537*584mm (W * D * H)
Weight :50KG
Warranty :1 Year
Measuring range :0-100%
Power supply :AC 220V, 50/60Hz, 5A
Keyword :Minimum Oxygen Concentration Index Tester
Display :Touch screen
Combustion tube height :500mm
Combustion tube flow rate :(40 ± 2) mm/s
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

Introduction:
This Minimum Oxygen Index Test Apparatus measures minimum oxygen concentration as necessary when specimen combustion.
The oxygen sensor is adapted Paramagnetic Type which has a precise, long life, less error rate and high reliability.

The Minimum Oxygen Index Test Apparatus meets the requirements of standards like ASTM D2863, ISO 4589-2, NES714 and so on.

Features:
1. Oxygen sensor, accuracy<0.1%;
2. Touch screen setting of oxygen percentage concentration, automatic adjustment;
3. Digitally display nitrogen flow rate and mixed gas flow rate;
3. One click setting, automatic ratio display of oxygen percentage concentration, without the need for manual adjustment and calculation;
4. High temperature and heat-resistant quartz glass cylinder, capable of withstanding high temperatures during testing;
5. Gas ignition device, with adjustable flame length and the ability to cut off the gas source;
6. Provide sample holders for supportable samples and unsupportable samples;
7. The quartz cylinder can automatically rise or fall, making it easy to replace samples;
8. Gas pipeline optimization design, providing faster response time;
9. The air circuit is a mass flow controller;
10.The controller adopts an imported programmable controller.
11.Adopt imported Paramagnetic oxygen analyzer and Mass flow meter.

Technical Specifications
1. Overal dimension: 400*537*584mm (W * D * H)
2. Oxygen concentration output: 0% -100%, set directly on the touch screen
3. Oxygen concentration measurement: 0-100%
4. Measurement resolution: 0.1
5. Measurement range: 0-100%
6. Working pressure: 0.05-0.3 Mpa.
7. Combustion tube height: 500mm
8. Inner diameter of combustion cylinder: 76mm
9. Restriction hole at the top of the combustion cylinder: Φ 40mm
10. Combustion tube flow rate: (40 ± 2) mm/s
11. Response time 1 second
12. Measurement accuracy: ± 0.2% (10% -60%) accuracy within the range
13. Test time: 0.1s~999.9s
14. Residual flame recording time: 0.01s~999.9s
15. Ignitioin method: manual ignition
16. Box material: aluminum and stainless steel
17. Sample holder: multiple options available
18. Glass cylinder: high-temperature resistant quartz glass cylinder with a diameter of 75mm and a mouth of 40mm;
19. Igniter: There are two methods: handheld igniter and machine controlled igniter (propane);
20. Oxygen control: mass flow meter controller
Nitrogen control: mass flow meter controller